Camille Bauer LINAX A325 Operating Instructions Manual page 47

Table of Contents

Advertisement

Advanced Graphic Data Manager
"System" menu
items
Analysis 2
Analysis 3
Analysis 4
Synchron. time
Week starting on
Only if "Weekly analysis" is
activated
Reset to zero
Submenu "Autom. printout"
Configurable parameters
(factory settings are highlighted in bold)
Here, choose whether a daily, weekly, monthly or annual analysis should
be performed.
"No": no analysis
"Externally controlled": the analysis is started or stopped by means of a
digital input (set the digital input to "Control input" and the action to
"Analysis x").
"Daily analysis": determines the daily minimum, maximum and average
values, and the volumes for the day, when the day changes.
"Weekly analysis": determines the weekly minimum, maximum and
average values, and the volumes for the week, when the week changes.
"Monthly analysis": determines the monthly minimum, maximum and
average values, and the volumes for the month, when the month
changes.
"Annual analysis": determines the annual minimum, maximum and
average values, and the volumes for the year, when the year changes.
Time for ending the signal analysis. Example for daily analysis: starting at
the set time, e.g. 7 a.m., the daily analysis continues until 7 a.m. the
following day. It contains the analysis for the measured values over the
last 24 hours. Factory setting: "00:00"
Specify what day the weekly analysis should start. Factory setting:
"Monday"
The analyzed values can be reset to zero. Example: reset after
commissioning a plant. All commissioning signals are discarded. The
graphics or memory storage are not influenced (traceability)!
Picklist: no, analysis 1-4, totalizer, all counters.
Note: All previous (commissioning) signals are discarded. The graphics/
storage are, however, not affected if you leave the setup and answer the
"Accept setup?" question with "No" (traceability). If you answer the
question with "Yes", the contents of the memory and thereby the graphic
display is also deleted/reset. If the previous signals are still required, save
them onto an external storage medium (USB stick or SD card) before
resetting. The reset takes effect immediately if you confirm acceptance
here.
Specify whether the analysis should be printed out automatically at the
end of an analysis.
Note: The information is only printed out if a USB printer is connected
to the device!
With the Batch option, the printout process is configured in the Batch
mode/printout menu.
Analysis 1: picklist: no, yes
Analysis 2: picklist: no, yes
Analysis 3: picklist: no, yes
Analysis 4: picklist: no, yes
Commissioning and use during operation
Direct access
17005/000
17010/000
17015/000
17020/000
17025/000
17035/000
17100/000
17105/000
17110/000
17115/000
code
47