Commissioning and use during operation
"System" menu
items
Submenu "Signal
Setup in order to get signal analysis for a preset time range/cycle as well as the function to manually reset the
analysis"
signal analysis. Up to 4 analyses can be performed in parallel. The analyses can be displayed in the Extras
menu.
Fig. 30:
Analysis 1
46
Configurable parameters
(factory settings are highlighted in bold)
IP address
Enter the IP address for the unit here. This IP address is assigned by your
network administrator. Please talk to your network administrator about
this. The unit is delivered with a preset IP address which must, however,
be changed in the setup procedure. Before you can make the input in the
unit, an IP address valid for your network must be defined.
Note: Entry only needed if the IP address is not acquired automatically
via DHCP.
The IP address must be unique within the network! Please be aware that
this number is not arbitrary, rather it must be defined dependent on the
network address of the TCP/IP network. The input form corresponds to
syntax (e.g. 192.168.100.002).
Subnetmask
Please input the subnetmask (given by your network administrator). The
subnetmask must be input if the unit is to establish connections in
another sub-network. Specify the subnetmask of the sub-network in
which the unit is located (e.g. 255.255.255.000).
Note: the IP address determines the network class. The result of this is
a default subnetmask (e.g. 255.255.000.000 for a Class B network).
Gateway
Please input the Gateway (given by your network administrator). Enter
the IP address of the gateway here if connections are to be established in
other networks. Factory setting: 000.000.000.000
Note: Changes to the system parameters are not activated until the
System menu is closed and the settings accepted. Only then will the unit
work with the new settings.
Setup - System, submenu "Signal analysis"
Determines the minimum, maximum and average values, as well as the
volumes and operating times, for the set time range.
"Externally controlled": the analysis is started or stopped by means of a
digital input (set the digital input to "Control input" and the action to
"Analysis x").
Picklist: no, externally controlled, 1 min to 12 h