Camille Bauer LINAX A325 Operating Instructions Manual page 24

Table of Contents

Advertisement

Operation
24
Operating
Operating function
element (Item
(Display mode = measured value display)
No.)
(Setup mode = operating in the Setup menu)
8
In the Display mode: alternating display indicating what percentage of the SD card or USB stick has
already been written to.
Status symbols are displayed for the following functions (alternate with the memory information):
simulation mode, data storage active, operating lock, batch active
In the Setup mode: the current "direct access" operating code is displayed
9
In the Display mode: screen for measured value display
Displays the current measured values, and the status in a fault/alarm condition, depending on the
signal display selected. In the case of counters, the type of counter is displayed as a symbol
If a measuring point has limit value status, the corresponding channel identifier is highlighted in red
(quick detection of limit values). When you are operating the unit, measured value acquisition
continues to run without interruption.
10
In the Display mode: alternating status display (e.g. set zoom range) of the analog or digital inputs in
the appropriate color of the channel.
In the Setup mode: different information can be displayed here depending on the display type.
a
Slot for SD card
Risk of data loss!
Do not remove the SD card if the yellow LED (d) is lit!
b
USB B socket type "function" e.g. for laptop
c
USB A socket type "host" e.g. for USB stick
d
LED at SD slot
Yellow LED lit when the unit writes to the SD card or reads it.
Risk of data loss!
Do not remove the SD card if the yellow LED (d) is lit!
Overview of symbols, see  Kap. 5.4
1)
Advanced Graphic Data Manager
1)
1)
.

Advertisement

Table of Contents
loading

Table of Contents