Commissioning and use during operation
90
Fig. 67:
Extras "Search in trace" - Search results
Subsequent reporting
Under "Store text", comments can be entered on the specific time selected by pressing softkey 4
"Extras". The date and time are automatically taken from the search result. A predefined text can
be selected or a new text can be entered (see "Setup –> Application –> Text). These texts are stored
in the event log/audit trail.
When user administration is activated ("Setup –> System –> Security –> Protected by:
"FDA 21 CFR Part 11"), this function is only available if a user is logged on and this user
has the necessary rights.
Zoom
In the history display, a larger time range can be configured by pressing softkey 4 "Extras" under
"Zoom".
Picklist:
"1:1": Every measured value is displayed.
"1:n": Only every nth measured value is displayed.
Note:
• The zoom function is only available for the following display modes: Curves, Curve in ranges,
Waterfall, Waterfall in ranges.
• The zoom value has to be configured separately for every group and display mode.
• No interpolation is performed, nor is the mean value determined.
• If the value for "n" is large, this can result in extended loading times.
• The zoom value does not affect the process for storing the measured value.
Search result when "Searching by results":