Advanced Graphic Data Manager
"System" menu
items
Timeout sequences
Configurable parameters
(factory settings are highlighted in bold)
Specify after how many seconds an order sequence is aborted if the
necessary data are not read in.
User input: 10 to 180 seconds; factory setting: 30 s
6.4.2
Setup - Inputs
Settings for the analog and digital inputs, mathematics channels (option) and linearization.
Depending on the selected function, the unit's user interface adapts itself, so that each time
only parameters that are required for safe unit functioning have to be checked/set.
Fig. 35:
Setup - Inputs
Setup - Inputs, submenu: Universal inputs
Procedure for the signal settings of the universal inputs:
Set the signal for the universal input (current, voltage, resistance thermometer, thermocouple, pulse, frequency)
Set the range for the universal input (e.g. 0-20 mA for current; not necessary for pulse/frequency)
Enter the measuring range and zoom range (not necessary for pulse)
Enter the damping/filter (not necessary for pulse/frequency)
Commissioning and use during operation
Start
?
Enter the channel name
Select the plot type
Enter the unit/dimension
?
Enter the number of decimal places
Direct access
code
10960/000