Commissioning and use during operation
"Inputs" menu
items
Upper frequency
Only for Signal "Frequency
input"
Meas. range end
Zoom start
Zoom end
Damping/filter
Comparison point
Only for Signal "Thermocouples"
Comparison temp.
Only for "Comparison point -
external"
Totalizer
Only for Signal "Pulse counter"
54
Configurable parameters
(factory settings are highlighted in bold)
Specify the upper frequency that corresponds to the end of the
measuring range. A maximum frequency of 12.5 kHz can be configured
here. Factory setting: 1000 Hz
The same as "Range start". Enter the end value for the measurement
range here, e.g. "14" for a 0-14 pH transmitter.
Note: Cannot be changed for resistance thermometers and
thermocouples!
If the total transmitter range is not being used then the lower value of
the required range can be entered here (higher resolution). Example:
transmitter 0-14 pH, required range: 5-9 pH. Set up "5" here. The zoom
has no influence on the storage.
As in "Zoom start". Enter the upper value of the required range here.
Example: transmitter 0-14 pH, required range: 5-9 pH. Enter "9" here.
The more unwanted interference there is on the measurement signal the
higher the value that should be entered here. Result: fast changes will be
damped/suppressed. Factory setting: 0.0 s
Only used when directly connecting thermocouples.
"Internal": compensation of the voltage error by measuring the
terminal temperature.
"External": compensation of error voltage by using thermostated
comparison points.
Note: When using thermocouples (TC) with an internal compensation
point and storage cycle of 100 ms the temperature of the compensation
pointing is measured every 5 seconds and the measurement is frozen for
1200 ms by the measurement switchover. This behavior appears only at
the above mentioned set-up and is visible due to the fast 100 ms storage
cycle. E.g. this can be avoided by the use of an external compensation
point.
Settings for the external comparison temperature (only when connecting
thermocouples directly).
Initial setting for the totalizer. Useful when continuing measurements
recorded to date with an (electro)-mechanical counter.
Factory setting: 0
Advanced Graphic Data Manager
Direct access
code
20075/000
to
20075/039
20080/000
to
20080/039
20085/000
to
20085/039
20090/000
to
20090/039
20095/000
to
20095/039
20100/000
to
20100/039
20105/000
to
20105/039
20300/000
to
20300/039