Advanced Graphic Data Manager
"System" menu
items
Submenu "Messages"
Settings for displaying/confirming messages. For example, messages can be:
(only available in the
– Messages triggered by set points
Expert setup)
– Messages triggered by a digital input
– Error messages, etc.
Fig. 32:
Acknowledging messages
Double messages
Can only be configured if "Do
not save" is selected for
"Acknowledging messages"
Switches relay
Submenu
To increase the life span of the LCD the rear illumination can be switched off (= screen saver).
"Screensaver"
If the unit is in an alarm condition (e.g. open circuit or alarm point violation), the screen saver is not switched
on or is automatically switched off. In this way, alarms can be read even if the screen saver is activated.
Fig. 33:
Configurable parameters
(factory settings are highlighted in bold)
Expert setup - System, submenu "Messages"
The time the message is acknowledged can be saved in the events list.
Note: this setting cannot be changed if the user administration function
is activated (FDA 21 CFR Part 11).
Picklist: do not save, save
The unit can merge the same messages into one single message (with the
time stamp of the last message).
Note: this setting cannot be changed if the user administration function
is activated (FDA 21 CFR Part 11).
Picklist: combine, display
A relay can be switched as soon as a message that has to be
acknowledged (e.g. on/off messages, device errors, etc.) is displayed.
Picklist: not used, relay x (xx-xx)
Note: If there are several reasons for triggering a relay (e.g. 2 different
limit values), this must be set to "Collective relay" under "Setup ->
Outputs -> Relay -> Relay x".
Setup - System, submenu "Screensaver"
Commissioning and use during operation
Direct access
19005/000
19000/000
19010/000
code
49