Commissioning and use during operation
"System" menu
items
Submenu "External
Settings for the external data carrier, amongst other things which data are to be stored in which format on the
memory"
external data carrier.
Fig. 31:
Save as
Memory buildup
Separator for CSV
Can only be configured for
"Open format (*.csv)"
Date/time
Can only be configured for
"Open format (*.csv)"
Operating time
Can only be configured for
"Open format (*.csv)"
Warning at
Can only be configured for
"Stack memory"
Switches relay
Can only be configured for
"Stack memory"
48
Configurable parameters
(factory settings are highlighted in bold)
Setup - System, submenu "External memory"
"Protected format": all the data are stored in a way that protects them
from data manipulation. They can only be interpreted using the PC
software supplied.
"Open format": the data are stored in CSV format. This can be opened by
a number of different programs (caution: no manipulation protection).
MS Excel restricts the open format to a maximum of 65535 rows.
"Stack memory": no more data can be stored once the data carrier is
full.
"Ring memory (FIFO)": once the data carrier is full the oldest data are
deleted so that new data can be stored (first in first out (FIFO)). Option
only possible for "Protected format".
Specify which separator is used by your application (e.g. in Excel =
semicolon). Picklist: comma, semicolon
Specify whether the date and the time should be saved in a joint column
or in two separate columns when saving the data in CSV format.
Specify the format for which operating times should be saved/displayed.
Factory setting: 0000h00:00
Gives a warning before the data carrier is x% full. A warning is indicated
on the unit and this is also stored in the event list. A relay can also be
switched.
Note: only with the external SD card (does not apply for USB stick)!
Factory setting: 90%
When the warning "Data carrier full" is displayed a relay can also be
switched.
Picklist: not used, relay x (xx-xx)
Note: If there are several reasons for triggering a relay (e.g. 2 different
limit values), this must be set to "Collective relay" under "Setup ->
Outputs -> Relay -> Relay x".
Advanced Graphic Data Manager
Direct access
code
12005/000
12000/000
12010/000
12011/000
12015/000
12020/000
12025/000