4.12.2

LV-DIC/LV-DIHC slider (Senarmont method)

Attach the DIC slider in Senarmont method to the D-ND6 revolving nosepiece.
DIC slider
4.12.3
D-DA analyzer slider, D-LP λ plate
Attach the D-DA analyzer and the D-LP λ plate to the D-ND6 revolving nosepiece.
Sextuple nosepiece
Chapter 4 Assembly
DIC slider slot
Sextuple nosepiece
Analyzer slider
λ plate
1 Select the slider (A or B) corresponding to the
objective to be used (refer to 3.16)
2 Insert the slider into the slot corresponding to the
position of the objective. Push the slider into the
limit to place the DIC prism in the optical path.
3 If you do not want to use the DIC slider, remove it
from the revolving nosepiece.
Two slots are provided for the revolving nosepiece.
Insert the analyzer slider into the slot (indicated with
"A") far from the objective, and insert the λ plate into
the slot (indicated with "λ") near the objective.
Insert these two sliders with their nameplates down.
When the sliders hit the second click-stop position,
the analyzer or D-LP λ plate enters the optical path.
Pull them out to the first click-stop position, and the
dummy hole enters the optical path.
* These sliders are designed to perform the
simplified polarization microscopy under the
diascopic illumination. Insert the dummy slider into
the slot for other microscopy as these sliders are
not required.
84