Attaching the DS-L2
Mount arm
attaching/detaching knob
Cable connection between the DS-L2 and MA200
(1)
(3)
(2)
Chapter 4 Assembly
1 Remove the one-touch release attached to the
front of the mount arm for the supporting arm.
Pull up the mount arm attaching/detaching knob,
One-touch
then hold up the docking section forward to
release
remove it.
2 Detach the stand arm attached on the rear of the
Rotation torque
nuts (x2)
DS-L2 by removing the four clamp screws.
3 Fix the docking section of the one-touch release to
Docking
the rear of the DS-L2 by using the clamp screws
section
(4) removed in Step 2.
(both sides)
4 Insert the docking section into the upper
one-touch release. Align the pins on the mount
arm with the holes on the docking section to insert
Supporting arm
it.
The DS-L2 attached to the supporting arm can be
removed by pulling up the mount arm
attaching/detaching knob.
6 The rotation torque can be adjusted depending on
the tightness of the docking section of the mount
arm to keep the DS-L2 steady, while achieving
more comfortable operation for angle adjustment.
Turn the two nuts on the docking sections with the
supplied tool and control each rotation torque for
the horizontal/vertical direction.
1 Connect the DS camera attached to MA200 with
the Camera connector on the left-side of the
DS-L2 using the camera cable exclusive for the
DS camera.
(2)
2 Connect the USB (H) connector on the right-side
of the DS-L2 and the [USB] connector on the rear
of the MA200 main body using the USB cable.
3 Connect the AC adapter exclusive for the DS-L2 to
the 12 VDC IN connector on the left side of the
DS-L2.