Id Circular Graduated Stage For Polarizing Microscope; Ml Use Of Optional Accessories (Sold Separately); P-Cs Senarmont Compensator - Nikon ECLIPSE E600 POL Instructions Manual

Polarizing microscope
Table of Contents

Advertisement

3. Operation Of Each Part
I
llJ
Circular graduated stage for polarizing microscope
Stage rotation
Loosening the stage rotation clamp screw can rotate the
stage. The angle of rotation can be read to 0.1 degrees
from the two vernier scales.
45° click-stop
Turning the 45° click-stop lever towards the front until it
stops causes the click-stop to move in 45° increments
from the position where the lever was rotated. This
makes it possible to move easily and accurately from the
extinction position to the diagonal position. Turning the
lever towards the back until it stops releases the click-
stop device.
Stage rotation
clamp
screw
Release the click-stop device where the stage has dropped into a click position. If the click-
stop device is released where the stage is not at this position. the first 45° increment will
not be accurate the next time the click-stop device is operated.
m
Use of optional accessories (Sold Separately)
P-CS Senarmont Compensator
The P-CS Senarmont compensator is used by inserting
into the slot of the intermediate tube in place of the
P-CL 1141.
&
tint plate.
It can be used to measure retardation up to I A according
to the following procedure.
It is normal to bring the top lens of the condenser into the optical path when observing at a
magnification of
!Ox
or greater. Measuring the retardation or evaluating by interference
color. however. stop down the condenser aperture diaphragm or swing out the top lens even
for a magnification of
!Ox
or higher (with the aperture diaphragm fully opened). and
illuminate with light that is as parallel to the optical axis as possible.