Condenser Refocusing Clamp; Condenser Mount Rotation - Nikon ECLIPSE MA200 Instructions Manual

Inverted metallurgical microscope
Table of Contents

Advertisement

Filter
ND2
(Neutral Density
filter)
ND16
(Neutral Density
filter)
GIF
(Green Interference
filter)
NCB11
(Neutral Color
Balancing filter)
3.23.6

Condenser refocusing clamp

Condenser
refocusing
clamp
Condenser
focus knob
Figure 3.23-7 Condenser refocusing clamp
3.23.7

Condenser mount rotation

Condenser mount rotation
clamp screw
Figure 3.23-8 Condenser mount rotation
Chapter 3 Operation Details
Adjusts the brightness for general microscopy or photomicroscopy. Reduces the light
intensity to 1/2. Dims the quantity of light to 1/2 (Transmittance: approx. 50%).
Adjusts the brightness for general microscopy or photomicroscopy. Dims the quantity
of light to 1/16 (Transmittance: approx. 6.3%).
Improves the contrast for microscopy under a monochrome light or when performing
monochrome imaging.
Corrects the color temperature for general microscopy or filming by daylight color TV
camera photography. Color reproduction is best when this filter is inserted into the
optical path, and the lamp voltage is set to the same display as the lamp rating.
Remove from optical path when performing monochrome imaging.
Condenser
focus knob
Condenser mount
Filter list
Description
Form the field diaphragm image on the sample
surface by turning the condenser focus knob. Turn
the condenser refocusing clamp clockwise to the limit
to mark this position.
When the condenser is elevated to change the
sample, it can easily be brought back down to the
initial position (at which the field diaphragm image is
formed) by turning the condenser focus knob to the
limit. The condenser refocusing clamp has a range of
motion of 13 mm.
The condenser mount can be turned by loosening the
condenser mount rotation clamp screw.
61