3.25

Using the DS Camera Control Unit (DS-L2)

Once you connect the DS Camera Control Unit DS-L2 with the DS camera attached to the vertical tube on
MA200 or the back port using the specified cable, you can operate the DS camera. Additionally, by
connecting MA200 with DS-L2 via USB cable, the following functions are available on the DS-L2.

• Displaying the microscope info menu

• Measurement corresponding to the magnification value of the microscope
• Recording the microscope info while saving the image
3.25.1

Procedure for turning on the power

1 Check that the MA200 main body and the DS-L2 are connected correctly. (Refer to "Cable Connection for
the DS-L2 and MA200" in 4.16.)
2 Set the power switch on the MA200 main body to the "|" side to turn it on.
3 Push in the power switch on the DS-L2 to turn it on.
3.25.2
Displaying the microscope info menu
Right-click on the mouse over the upper-screen of the DS-L2 to display the microscope info menu. (Refer to
Figure 3.25-1.)
Light source for the episcopic illumination
(EpiLamp) and brightness value (lamp voltage)
(If the motorized C-HGFIE HG Precentered Fiber
Illuminator is used as the light source for the
episcopic illumination (EpiLamp), the transmittance
of the ND filter for the fiber illuminator is indicated in
the brightness value area.)
Figure 3.25-1 DS-L2 microscope info menu
Chapter 3 Operation Details
Magnification/NA value
of the objective
Magnification value of
the intermediate
magnification unit
The microscope info menu buttons
: Closes the microscope info menu.
: Resizes the microscope info menu.
: Displays the objective info. (Refer to 3.25-2.)
To display the objective lens info on the DS-L2,
use the "MA200 Setup" in "MA200 Setup Tool"
supplied with MA200 to allow the user to
register the info from an external PC to the
microscope. For installing and operating
"MA200 Setup," refer to the software manual of
"MA200 Setup Tool."
to
Specify the nosepiece address to be placed into
the optical path. This button appears when the
motorized LV-NU5A is attached to MA200.
However, this is not applied to the manual
nosepiece. Press the button, then the revolving
nosepiece is automatically set to the nosepiece
address. The nosepiece address that is
currently placed into the optical path is framed in
red.