4.18

Camera Adapter

You can perform photomicroscopy by mounting the photomicrographic equipment such as CCD camera on
the microscope with the LV-TV TV or the C-mount adapter attached.
Apply the camera adapter to the vertical tube port on the MA2-TI3 trinocular eyepiece or the back port on the
upper-surface of the microscope. Be sure to apply the adapter or zoom lens compatible with the camera.
Clamp
screw
CAUTION
After mounting the camera, be sure to place cables into the cable keeper. If a cable comes into contact
with the lamphouse, the cable sheath may melt resulting in an electrical shock or fire.
4.19

Power Cord

WARNING
Make sure to use the specified power cord. Using a wrong power cord may result in malfunctions or
fire. This product is classified as subject to Class I protection against electrical shock. Make sure it is
connected to an appropriate ground terminal (protective earth terminal).
For specifications of the power cord, refer to "7. Specifications."
Turn off the power switch on the product (set the switch to the "O" side).
Insert the plug of the power cord into the AC inlet on the back of the product. Then, securely plug the power
cord to the outlet.
Chapter 4 Assembly
Camera
1 Remove the cap from the vertical tube port/back
port.
2 Fully loosen the clamp screw of the side of the
Camera adapter
vertical tube port/back port.
3 Insert the camera adapter into the hole on the
vertical tube port/back port and secure it by
tightening the clamp screw.
Clamp
screw
97