Nikon ECLIPSE MA200 Instructions Manual page 95

Inverted metallurgical microscope
Table of Contents

Advertisement

Attach the T-P2 polarizer slider.
Fixing ring
Pin
Groove
Condenser
mount
Rotation clamp
screw
White
index mark
Mount rotation
clamp screw
Chapter 4 Assembly
Steel ball
and groove
Polarizer
rotation lever
White
index mark
1 Remove the system condenser, if attached, from
the condenser holder to attach the T-P2 polarizer
slider.
2 Mount the polarizer slider on the condenser
mount, and align the positioning pin for the
polarizer slider with the groove on the condenser
mount. Turn the fixing ring to secure the polarizer
slider.
3 Loosen the rotation clamp screw, hold the
polarizer rotation lever, and turn the rotating
section to align the white index mark with the
center of the scale plate. (Be sure that the steel
ball on the tip of the clamp screw is in the groove
on the rotating section.) Tighten the rotation clamp
screw to secure the rotating section.
4 Use a hexagonal screwdriver to loosen the
condenser mount rotation clamp screw on the left
side of the condenser holder. Turn the entire
polarizer to align the white index mark set in Step
3. with the white index mark on the condenser
holder. Tighten the mount rotation clamp screw to
secure the mount.
5 The polarizer can now be moved in and out of the
optical path by moving the slider.
93