Using The Hg Precentered Fiber Illuminator (C-Hgfi/C-Hgfie); Procedure For Turning On The Power Switch - Nikon ECLIPSE MA200 Instructions Manual

Inverted metallurgical microscope
Table of Contents

Advertisement

3.22

Using the HG Precentered Fiber Illuminator (C-HGFI/C-HGFIE)

Figure 3.22-1 HG Precentered Fiber Illuminator
3.22.1

Procedure for turning on the power switch

LAMP indicator
POWER indicator
Figure 3.22-2 Procedure for turning on
the power switch
Chapter 3 Operation Details
The specified light source LV-LH50PC may not
provide sufficient brightness to perform epi-fl
microscopy. In such case, use an HG precentered
fiber illuminator that uses a high intensity mercury
lamp as a light source.
HG Precentered
Fiber Illuminator
1 Be sure that the fiber illuminator is properly
2 Set the POWER switch to the "|" side to turn on
2
Power switch ON
3 Set the POWER switch to the "|" side to turn on
4 To turn off the fiber illuminator, set the power
* To use the HG precentered fiber illuminator
1
Power switch ON
56
connected to the MA200 main body.
the fiber illuminator and wait until it starts up.
When the POWER indicator on the fiber
illuminator lights up, the LAMP indicator starts
blinking. Once the illuminator starts up, the LAMP
indicator stops blinking and lights up.
If the C-HGFIE motorized fiber illuminator is in
use, the communication function on the illuminator
does not activate until it starts. Be sure to wait until
the illuminator starts up.
the power on the MA200 main body.
The power LED on the front of the MA200 main
body lights up.
switch to the " " side in the sequence of the fiber
illuminator, the MA200 main body.
correctly, carefully read the instruction manual
supplied with the fiber illuminator and make sure
to follow the instruction.