Attach the filter and the slider.
• Do not touch filters or other optical components with your bare hands.
• Latches at both ends of the filter slider are at their end point when sliding. When removing a filter
slider, you can slide it out by pushing up the latch on the opposite side of the tab with your finger to
release the filter slider. Applying undue force on the filter slider can break the latches.
Filter slider
(back)
Tab
Connect the TI-PS100W power supply.
EXTERNAL/CTRL
switch
AC inlet
Chapter 4 Assembly
Filter
Latches
POWER
switch
12 VDC/6 VDC
output connector
EXTERNAL/CTRL
connector*
1 Attach the desired filter to the filter slider.
Attach it from the back of the filter slider. The
mounting hole has three tabs to keep the filter
from falling. Only one tab can be moved to the
side. Move this tab aside, and attach the filter.
2 Affix a label indicating the filter type on the tab of
the filter slider.
3 Insert the filter slider into the slot of the supporting
pillar for dia-Illuminator.
The filter slider has latches that determine the limit
of slide operation. Push the latches up, and push
the filter into the slot.
You can insert up to four filter sliders.
You can insert filter sliders from the right or the
left. If they are all inserted from the same direction,
they will be difficult to handle, so you should insert
them alternately from the left and right.
Connect the cable of the 100W diascopic illumination
lamp to the 12 VDC output connector on the power
supply. Secure the lamp cable with the lock ring.
Do not connect the lamp cable of 100W diascopic
illumination lamp to the lamp connector [LAMP
DC12V50W] on the MA200 main body. It is possible
to connect the cable to the connector because the
type of the cable connector is the same as the one on
the lamp connector. However, if you turn the
brightness control dial to the maximum, the light turns
dull being caused by inappropriate connection.
Connect the lamp cable of 100W diascopic
illumination lamp to the TI-PS100W power supply as
described above.
* Do not use the EXTERNAL/CTRL connector.