Registering Information On The Objective; Procedure For Turning On The Power Switch - Nikon ECLIPSE MA200 Instructions Manual

Inverted metallurgical microscope
Table of Contents

Advertisement

4.3.3

Registering information on the objective

You can register the information to be checked on a PC or the DS-L2 with the "MA200 Setup" in the supplied
"MA200 Setup Tool" software. For details on how to install the "MA200 Setup" on an appropriate device, refer
to the software manual supplied with "MA200 Setup Tool."
4.3.4

Procedure for turning on the power switch

To establish an appropriate connection between MA200 and an external device or to use the product safely,
turn on the power as follows.
■ For using the C-HGFIE HG Precentered Fiber Illuminator (motorized type)
1 Set the power switch on the fiber illuminator to the "|" side and wait until it starts up.
2 Set the power switch on the MA200 main body to the "|" side to turn it on.
3 To turn off the fiber illuminator, set the power switch to the " " side in the sequence of the fiber illuminator,
the MA200 main body.
■ For using the Nosepiece Controller 2
1 Set the power switch on the Nosepiece Controller 2 to the "|" side to turn it on.
2 Set the power switch on the MA200 main body to the "|" side to turn it on.
■ For using the C-HGFIE HG Precentered Fiber Illuminator (motorized type) and the Nosepiece Controller 2
1 Set the power switch on the fiber illuminator to the "|" side and wait until it starts up.
2 Set the power switch on the Nosepiece Controller 2 to the "|" side to turn it on.
3 Set the power switch on the MA200 main body to the "|" side to turn it on.
4 To turn off the dia pillar illuminator, set the power switch to the " " side in the sequence of the fiber
illuminator, the MA200 main body, the Nosepiece Controller 2.
■ For using the DS-L2/DS-U2
1 Set the power switch on the MA200 main body to the "|" side to turn it on.
2 Push in the power switch on the DS-L2 or DS-U2 to turn it on.
■ For using the MA2-DP Supporting Pillar for Dia-Illuminator 100W under the diascopic illumination
1 Set the brightness control dial on the MA200 main body to OFF.
2 Set the power switch on the power supply (TI-PS100W) to the "|" side to turn it on.
3 Set the power switch on the MA200 main body to the "|" side to turn it on.
Chapter 4 Assembly
76

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents