Supporting Arm (for DS-L2)

You can mount the camera control unit DS-L2 on the MA200 main body by attaching the supporting arm.
Chapter 4 Assembly
1 Loosen clamp screws (2) using the supplied
hexagonal screwdriver and remove the cover from
the attaching part of the supporting pillar for
dia-Illuminator 100W/supporting arm on
top-surface of the microscope.
2 Align the pins (2) on the lower-surface of the
supporting arm with the holes on the attaching
part. Using the supplied hexagonal wrench (5
mm), secure the supporting arm by tightening the
four hexagonal socket head bolts (M6 x 20 mm).
The standard descriptions on the DS-L2 attaching
part is shown as follows.
• VESA 75 mm standard compliance,
UL1678-compatible
• Attachment screw holes: M4 depth 7 mm
• Tightening torque: 80 to 120N cm
• Leave a clearance of about 100 mm around the
DS-L2 to prevent heat from collecting near the
equipment.