Tests and Adjustments
PNA Series Microwave Network Analyzers
ANSI/NCSL Z540–1–1994 Verification
E8362C, E8363C, E836C
ANSI/NCSL Z540–1–1994 Verification
To meet the criteria for ANSI/NCSL Z540-1-1994, perform the preliminary checks and all performance tests
1
without stopping to repair or adjust
. Refer to
Figure 3-1
for test flow. Print data at the completion of all the
tests, even if you are aware that the analyzer did not pass. If there is a failure, complete the verification
before you troubleshoot, repair, and adjust. After the failure has been corrected, repeat the entire set of
performance tests and generate a new set of data.
Figure 3-1
ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
1. Stop only in case of a catastrophic failure or cable connector damage.