Agilent Technologies E8362C Service Manual page 55

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362C, E8363C, E836C
Precautions for Handling Airlines
The verification kit includes airlines. Follow the steps below to ensure good measurement techniques.
• Be very careful not to drop the airline's center or outer conductor. Irreparable damage will result if these
devices are dropped.
• During this procedure, you will be touching the exposed center conductor of the test port and the center
conductor of the airline.
— Observe proper ESD procedures.
— Clean your hands or wear gloves; skin oils will cause a change in electrical performance.
• Refer to
Figure
3-9,
Figure
Figure 3-9
Aligning the Center Conductor
Figure 3-10
Torquing the Connection
Service Guide E8364-90038
3-10, and to your calibration kit manual for connection procedures.
Tests and Adjustments
System Verification
3-23