PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
Table 5-5
AUX I/O Connector Pin Assignments
Pin Numbers
Name
1
ACOM
2–3
Analog Out 2–1
4
No connect
5
DCOM
6–8
Pulse Out 3–1
9
+5 V
10
Pass/Fail Write Strobe
11
Sweep End
12
Pass/Fail
13
Output Port Write Strobe
14
Analog In
15
ACOM
16
Power Button In
17
DCOM
18
Ready for Trigger
19
External Trigger In
20
Footswitch In
21
+22 V
22–25
In/Out Port C0–3
Service Guide E8364-90038
DB-25 Male Connector
Function
0 V, ground reference for analog signals
10 to +10 Vdc output, 10 mA max, Ro = 100 W
For future enhancements
0 V, ground reference for digital signals
TTL out, programmable pulse (for future use)
+5 Vdc output, 100 mA max.
TTL out, active low (1 s min) indicates pass/fail line is valid
TTL out, active low (10 s min) indicates sweep is done
TTL out, latched, indicates pass or fail (programmable polarity)
TTL out, active low (10 s min) writes I/O port data
10 to +10 Vdc input, Ri = 100k W
0 V, ground reference for analog signals
Open-collector input, active low replicates a power button press
0 V, ground reference for digital signals
TTL out, low indicates ready for external trigger
TTL in, level trigger (width >1 ms), programmable polarity
TTL in, active low (width >1 ms), triggers a programmable event
+22 Vdc output, 100 mA max.
TTL I/O, general purpose
Theory of Operation
Source Group Operation