Agilent Technologies E8362C Service Manual page 50

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

Tests and Adjustments
PNA Series Microwave Network Analyzers
Preliminary Checks
E8362C, E8363C, E836C
Figure 3-7
Typical Cable Connector Repeatability Response
If the Cable Connector Repeatability Check Fails
1. Clean the cable and devices, and torque connections to specifications (refer to
"Review the Principles of
Connector Care" on page
3-5). Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.
3-18
Service Guide E8364-90038