Troubleshooting
Measurement System Troubleshooting

Verifying the A, B, R1, and R2 Traces (Standard S-Parameter Mode)

NOTE
There is no way to view the frequency offset receiver response (Option 080). However, some
standard S-parameter receiver trace information is helpful in troubleshooting the frequency
offset section of the PNA. It is therefore recommended that you run this test even if you only
suspect the frequency offset section of malfunctioning.
The first step is to verify that the A, B, R1, and R2 traces are present and that they are approximately level:
• Connect an Open or Short standard from a mechanical calibration kit to each test port (use adapters if
necessary).
• On the
menu, point to
System
• Traces A, B, R1, and R2 are displayed in four separate data windows as shown in
discrepancies of the traces in these windows can help you to isolate the faulty assembly.
Figure 4-6
Typical Four Channel Display
4-26
,
, and then click
Service
Utilities
PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
.
Receiver Display
Figure
Service Guide E8364-90038
4-6. Identifying