Initialize the Windows Operating System
1. Connect a mouse and keyboard to the analyzer.
2. Turn on power to the analyzer.
3. The operating system initialization process will run automatically. Part way through the process, you
will be prompted to accept the Windows licensing agreement. The entire process takes approximately
six minutes, during which the analyzer will reboot up to three times.
4. After the PNA application screen is displayed, there may be a display to remind you to upgrade the
firmware. If so, click OK to continue.
Set Up the Agilent Administrator Password
After an HDD replacement, it is necessary to set up the Agilent administrator password for the analyzer.
Go to the following Internet site, enter your PNA serial number, and get a six-digit numerical password:
http://na.tm.agilent.com/pna/hdd/index.html#agpswd.
The network analyzer must be powered up and operating for this procedure.
1. On the Windows Desktop, right-click on My Computer, and then click Manage.
2. In the Computer Management window, navigate to System Tools, Local Users and Groups, Users.
3. Right-click on Agilent, click on Set Password, and complete these steps in the window:
a. Type the password obtained from Agilent as the Password. Refer to
Administrator Password" on page 7-67
b. Type the password again in the Confirm password block.
c. Click OK.
4. Right-click on Agilent, click on Properties, and complete these steps in the window:
a. Unselect User must change password at next logon (if not already unselected).
b. Unselect User cannot change password (if not already unselected).
c. Select Password never expires (if not already selected).
d. Click OK.
5. Close the Computer Management window.
Install Backup Files onto the New Hard Disk Drive
The files that were previously saved onto a USB memory drive must now be installed onto the new HDD. The
network analyzer must be powered up and operating.
1. Insert the USB memory drive containing the backup files into a USB port.
2. Click on START. (It may be necessary to exit the Network Analyzer application.)
3. Point to Programs, Accessories, and then open Windows Explorer.
4. Navigate to C:\Program Files\Agilent\Network Analyzer.
5. Copy each of the backup files from the USB memory drive to the HDD.
Service Guide E8364-90038
Removing and Replacing the A41 Hard Disk Drive (HDD)
for information if you have not yet obtained this password.
Repair and Replacement Procedures
"Set Up the Agilent
7-67