3.8 Ghz Pmyo Frequency Adjustment - Agilent Technologies E8362C Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

Tests and Adjustments
Adjustments

3.8 GHz PMYO Frequency Adjustment

The 3.8 GHz adjustment is used to adjust the frequency accuracy of the network analyzer's 3.8 GHz
permanent-magnet YIG oscillator (PMYO) on the A12 source 20.
Equipment used for the Frequency Adjustment at 3.8 GHz
Equipment Type
Spectrum analyzer
RF Cable, 3.5 mm (f) to 3.5 mm (f)
Adapter, 3. 5 mm (m) to type-N (m)
RF Cable, 2.4 mm (f) to 2.4 mm (f)
Adapter, 2.4 mm (m) to type-N (m)
a. Must be capable of measuring a signal at 3.8 GHz.
Procedure
1. Connect the equipment as shown in
and the spectrum analyzer.
Figure 3-18
Setup for Adjustment of the 3.8 GHz PMYO
2. On the System menu, point to Service, Adjustments, and then click 3.8 GHz Freq Adjust.
3. Ensure the GPIB settings are correct.
4. Click Begin Adj, and then follow the instructions as they are displayed.
3-40
Models
Used With
All
E8362C
E8362C
E8363C, E8364C
E8363C, E8364C
Figure
3-18. Connect a GPIB cable between the network analyzer
PNA Series Microwave Network Analyzers
E8362C, E8363C, E836C
Model or
Alternate Model
Part Number
or Part Number
8565E
856xE
85131C
85131E
1250-1743
None
85133C
85133E
11903A
None
Service Guide E8364-90038
a

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8364cE8363c

Table of Contents