Removing The Covers - Agilent Technologies E8362C Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing the Covers

Removing the Covers
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
Removing the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front handles in the vertical
position. Do not place the analyzer on its front panel without the handles. This will damage
the front panel assemblies.
Refer to
Figure 7-1
for this procedure.
1. Disconnect the power cord.
2. With a T-20 TORX driver, remove the strap handles (item
ends until the handle is free of the analyzer.
3. With a T-20 TORX driver, remove the four rear panel feet (item
).
4. Slide the four bottom feet (item
5. Slide the cover off of the frame.
Removing the Inner Cover
Refer to
Figure 7-1
for this procedure.
1. With a T-10 TORX driver, remove the cover attachment screws (item
head and some are pan head. Note the locations of each type for reinstallation.
2. With a T-10 TORX driver, loosen the screws on the two retainer clips (item
3. Lift off the cover.
Replacement Procedure
Reverse the order of the removal procedures.
Be sure that the cover is located behind the two retainer clips (item
(item
). Remember to tighten the retainer clip screws. The retainer clips may rotate slightly when the
screws are tightened; this is normal and acceptable.
7-6
) off the cover.
PNA Series Microwave Network Analyzers
) by loosening the screws (item
) by removing the center screws (item
). Some of these screws are flat
).
) before installing the cover screws
Service Guide E8364-90038
E8362C, E8363C, E8364C
) on both

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8364cE8363c

Table of Contents