Repair and Replacement Procedures

Removing and Replacing the Lithium Battery

Removing and Replacing the Lithium Battery
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-37
for this procedure.
1. Disconnect the power cord.
2. Remove the A15 CPU board. Refer to
3. Lift the battery (item
). Be careful not to over bend the battery clip as it can be damaged or broken.
4. DO NOT THROW AWAY THE BATTERY. COLLECT IT AS SMALL CHEMICAL WASTE. Refer to
Battery Disposal" on page 1-9
Replacement Procedure
1. Reverse the order of the removal procedure following all instructions included with the new battery.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to
Table 7-2 on page
7-80
"Removing and Replacing the A15 CPU Board" on page
) just enough to free it from its holder and slide it from under the battery clip (item
for additional information on battery disposal.
7-82.