Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
E8362C, E8363C, E836C
8. Reconnect the power and USB cables. Turn on the analyzer.
9. On the System menu, point to Service, Adjustments, and then click LO Power Adjust.
10. Ensure the GPIB settings are correct.
11. Click Begin Adj, and then follow the instructions as they are displayed.
12. Turn off the analyzer, disconnect the power cable, reconnect the semi-rigid cables, and replace the
covers.
3-42