PNA Series Microwave Network Analyzers
E8362C, E8363C, E836C
4. If shorts and opens are not connected to all ports, you will be prompted to connect them as they are
needed.
5. The result of the operator's check will be shown as a PASS or FAIL next to each test (refer to
The PNA Operator's Check dialog box will look different for different PNA model numbers and installed
options. Some of the tests are performed only if the appropriate options are installed in the PNA.
Figure 3-3
Operator's Check Dialog Box
If the Operator's Check Fails
1. Clean the test ports, shorts, and adapters. Torque connections to specification. Repeat the check.
2. If the check still fails, suspect a faulty component. Refer to
page 4-24
to begin troubleshooting to determine the faulty component.
Service Guide E8364-90038
Tests and Adjustments
"Measurement System Troubleshooting" on
Preliminary Checks
Figure
3-3).
3-11