Ansi/Ncsl Z540-1-1994 Verification - Agilent Technologies N5230A Service Manual

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

Tests and Adjustments
PNA Series Microwave Network Analyzers
ANSI/NCSL Z540–1–1994 Verification
N5230A
ANSI/NCSL Z540–1–1994 Verification
To meet the criteria for ANSI/NCSL Z540-1-1994, perform the preliminary checks and all
1
performance tests without stopping to repair or adjust
. Refer to
Figure 3-1
for test flow.
Print data at the completion of all the tests, even if you are aware that the analyzer did not
pass. If there is a failure, complete the verification before you troubleshoot, repair, and
adjust. After the failure has been corrected, repeat the entire set of performance tests and
generate a new set of data.
Figure 3-1
ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
1. Stop only in case of a catastrophic failure or cable connector damage
3-8
Service Guide N5230-90014