Where To Begin Troubleshooting - Agilent Technologies E8362C Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
• If all traces are present and are similar to the traces in
the analyzer's measurement system. There may, however, be a minor failure in the analyzer.
To test further:
— Go to
Chapter 3 , "Tests and Adjustments,"
— If a problem still exists, contact Agilent. Refer to
• If any of the traces are not present, are noisy or distorted, or are at an incorrect level, then there is a
problem with the analyzer's measurement system. Proceed to

Where to Begin Troubleshooting

For the purposes of troubleshooting, the analyzer block diagram is divided into the following functional
groups:
• the source group
— A8 fractional-N synthesizer
— A10 frequency reference
— A11 phase lock
— A12 source 20
— A16 test set motherboard
— A17 LOMA 10
— A18 MA 20
— A19 splitter
— A20 LODA
— A21 SOMA 50 (E8363C and E8364C only)
• the signal separation group
— A22 switch splitter
— A23 and A24 detectors (E8363C and E8364C only)
— 30-dB fixed attenuators (E8362C only)
— A25 and A26 test port couplers
— A36 and A37 60-dB step attenuators (Option UNL only)
— A38 and A39 bias tees (Option UNL only)
— Port 1, Port 2, A, B, R1, and R2 ports
• the receiver group
— A27, A28, A29, and A30 first converters
— A31 receiver A
— A32 receiver R1
— A33 receiver R2
— A34 receiver B
— A35 receiver motherboard
— A6 SPAM
• the frequency offset group (Option 080 only)
— A9 fractional-N synthesizer (Option 080 only)
Service Guide E8364-90038
Measurement System Troubleshooting
Figure
4-6, then there are no major problems with
and perform all the tests in that section.
"Contacting Agilent" on page
"Where to Begin
Troubleshooting
2-12.
Troubleshooting".