Installing Or Changing A Serial Number - Agilent Technologies E8362C Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

General Purpose Maintenance Procedures
Option Enable Utility

Installing or Changing a Serial Number

It is necessary to reinstall the instrument serial number if the test set motherboard is replaced.
IMPORTANT Use extreme care when entering the serial number, as only one attempt is allowed.
1. To change a serial number, select
page
8-15). The current serial number is displayed. If no serial number has previously been entered, the
word "NONE" will be displayed.
2. VERY CAREFULLY, type the new serial number into the space provided and then click
3. If an error is made in entering the serial number, obtain a clear code from Agilent, enter the clear code in
the space provided, and then click
NOTE
To change an incorrect serial number, a clear-code password is required. Contact Agilent to
obtain the clear-code. Refer to
8-18
from the
Change Serial
Option Enable
. The correct serial number can then be entered.
Clear
"Contacting Agilent" on page
PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
menu bar (see
Figure 8-5 on
Change
2-12.
Service Guide E8364-90038
.

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8364cE8363c

Table of Contents