If Access Group (Option H11); A47 Rf Path Coupler; A48 Lo Path Coupler; A49 If Multiplexer Board - Agilent Technologies E8362C Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

Theory of Operation

IF Access Group (Option H11)

IF Access Group (Option H11)
The IF access group is additional hardware included in the PNA that allows for external connection to the
internal IF path of the analyzer.
The following assemblies and components are included in the IF access group:

A47 RF Path Coupler

A48 LO Path Coupler

A49 IF Multiplexer Board

Rear-Panel Connections

— TEST SET DRIVERS
— PULSE IN
— 8.33 MHz IF IN
A47 RF Path Coupler
The A47 RF path coupler is located in the RF path of the PNA between the A12 source and either the A21
SOMA 50 (for E8363/4C) or the A22 switch (for E8362C). A portion of the RF signal is coupled off and
delivered to the rear-panel TEST SET DRIVERS RF connector.
A48 LO Path Coupler
The A48 LO path coupler is located in the LO path of the PNA between the A18 MA24 and the A19 splitter. A
portion of the LO signal is coupled off and delivered to the rear-panel TEST SET DRIVERS LO connector.
A49 IF Multiplexer Board
The A49 IF multiplexer board provides the circuitry to allow the external pulse and IF signals to be input.
Rear-Panel Connections
TEST SET DRIVERS: A portion of the RF path signal and the LO path signal is coupled off and routed to the
rear panel TEST SET DRIVERS RF and LO connectors. These connectors provide access to the RF and LO
signals for use by an external test set.
PULSE IN: An RF pulse input connection is provided for each of the four receiver channels (A, R1, R2, and B)
via four rear-panel connectors. These signals are routed to the A49 IF multiplexer board where they can be
switched into the signal path of each of the four measurement channels.
8.33 MHz IF IN: An 8.33 MHz IF input connection is provided for each of the four receiver channels (A, R1, R2,
and B) via four rear-panel connectors. These signals are routed to the A49 IF multiplexer board where they
can be switched into the signal path of each of the four measurement channels.