Tests and Adjustments
PNA Series Microwave Network Analyzers
Preliminary Checks
E8362C, E8363C, E836C
Figure 3-6
Typical Cable Magnitude and Phase Stability Response
If the Cable Magnitude and Phase Stability Check Fails
1. Clean the cable and devices and torque connections to specifications (refer to
"Review the Principles of
Connector Care" on page
3-5). Repeat the check.
2. If the check still fails, the cable should be repaired or replaced.
3-16