Maintenance; Physical Maintenance; Electrical Maintenance; Analyzer Options Available - Agilent Technologies E8362C Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C

Maintenance

WARNING
To prevent electrical shock, disconnect the analyzer from the mains source before
cleaning. Use a dry cloth or one slightly dampened with water to clean the external case
parts. Do not attempt to clean internally.

Physical Maintenance

Clean the cabinet, including the front panel, using a dry or slightly damp cloth only.

Electrical Maintenance

Refer to
"Review the Principles of Connector Care" on page

Analyzer Options Available

Options as Upgrades

The options described in this section can be ordered as upgrades. Refer to
page
2-6.
Individual upgrade kits are also available and can be ordered by part number. Refer to
on page 6-85
and
"E8363C, E8364C Upgrade Kits" on page 6-85

Option 010, Time Domain

An Option 010 analyzer can display the time domain response of a network or test device by calculating the
inverse Fourier transform of the frequency domain response. This calculation allows the Option 010 analyzer
to show the response of a test device as a function of time or distance. In time domain, the magnitude and
location of a discontinuity and the characteristics of individual transmission paths of a network or test
device can be determined. Time domain operation retains all the accuracy of active error correction.

Option 014, Configurable Test Set

An Option 014 analyzer can be configured to measure high-power devices and devices for high dynamic
range.
For a high-power measurement, external amplifiers and high power attenuators or isolators can be added to
complete the test setup. Test port output power up to 1 Watt (+30 dBm) can be applied to the device under
test (DUT). Additionally, there is an external reference input that allows the external amplifier's frequency
response and drift to be ratioed out.
For high dynamic range measurements, front panel jumpers are moved to reverse the signal path through
one of the couplers, allowing for a 15 dB improvement in transmitted signal sensitivity in one direction only.
Jumpers are installed on both ports allowing the user to choose a measurement in either the forward or
reverse direction.
Service Guide E8364-90038
General Product Information
3-5.
"Analyzer Upgrades Available" on
"E8362C Upgrade Kits"
for a complete list of available kits.
Maintenance
2-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8364cE8363c

Table of Contents