Agilent Technologies E8362C Service Manual page 114

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
Checking the A8 Fractional-N Synthesizer Band 2 Output
Perform this procedure if you observe a problem in band 2 in all receivers.
1. Refer to the block diagram at the back of this chapter and to
page
6-16. Locate semirigid cable W19, at the A8 fractional-N synthesizer board.
CAUTION
Be careful not to damage the center pins of the semirigid cable. Some flexing of the cables is
necessary to measure the output.
2. Using a 5/16-inch torque wrench, disconnect W19 at A8J106.
3. Connect the spectrum analyzer to A8J106.
4. Set the network analyzer to a CW frequency of 1 GHz.
• The spectrum analyzer should measure a signal at 1.008 GHz.
5. If this signal is not present, replace the A8 fractional-N synthesizer board. Refer to
Replacing the A6, A8, A9, and A10 Boards" on page
6. If the signal is present, leave the spectrum analyzer connected to A8J106 and continue with
the A8 Fractional-N Synthesizer Band 3 through 25 Output" on page
Checking the A8 Fractional-N Synthesizer Band 3 through 25 Output
Perform this procedure if you observe a problem in bands 3 through 25 in all receivers.
1. Set the network analyzer to a CW frequency of 2 GHz.
• The spectrum analyzer should measure a signal at 2.008 GHz.
2. If the signal is not present, replace the A8 fractional-N synthesizer board. Refer to
Replacing the A6, A8, A9, and A10 Boards" on page
3. If the signal is present, reconnect cable W19, and then continue with
Bands (With All A8 Signals Present)" on page
Checking the A17 LOMA 10, All Bands (With All A8 Signals Present)
Perform this procedure if you observe a problem in any band and all signals from the A8 fractional-N
synthesizer are present.
1. Refer to the block diagram at the back of this chapter and to
page
6-24. Locate semirigid cable W18, at the A17 LOMA 10.
CAUTION
Be careful not to damage the center pins of the semirigid cable. Some flexing of the cables is
necessary to measure the output.
2. Using a 5/16-inch torque wrench, disconnect W18 at the A17 LOMA 10.
3. Connect the spectrum analyzer to the open A17 connector.
4. For each frequency band in which problems were seen: Set the network analyzer to each of the
frequencies indicated in the
page
4-36.
5. Set the spectrum analyzer to measure each of the corresponding frequencies in the column labeled:
"Expected A17 Output Frequency (GHz)"
4-34
7-18.
7-18.
4-34.
"Network Analyzer Source Frequency Setting (GHz)"
in
Table 7 on page
PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
"Top Cables–All, All Options Except 080" on
"Removing and
4-34.
"Removing and
"Checking the A17 LOMA 10, All
"Bottom Semi-Rigid Cables, No Options" on
column in
4-36.
Service Guide E8364-90038
"Checking
Table 7 on