PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
Figure 7-28
Hard Disk Drive Assembly Removal
Disassembling the HDDA
NOTE
If the HDD is being replaced by the HDDA (N8980A or N8981A), it is not necessary to
perform these steps.
Refer to
Figure 7-29
and
1. Remove the HDDA from the analyzer. Refer to
2. With a T-10 TORX driver, remove the following items that mount the HDD (item
a. The four screws (item
b. If the HDD is equipped with the serial ATA interface:
• the six bushings (item
• the nylon spacer (item
c. If the HDD is equipped with the parallel ATA interface:
• the eight bushings (item
3. Disconnect the HDD (item
Service Guide E8364-90038
Figure
7-30.
).
)
)
)
) from the interface board (item
Removing and Replacing the A41 Hard Disk Drive (HDD)
"Removal Procedure" on page
).
Repair and Replacement Procedures
7-62.
), to the bracket:
7-63