ANSI/NCSL Z540–1–1994 Verification
N5242A
ANSI/NCSL Z540–1–1994 Verification
To meet the criteria for ANSI/NCSL Z540-1-1994, perform the preliminary checks and
1
either system verification or performance tests without stopping to repair or adjust
. Refer
to
Figure 3-1
for test flow. Print data at the completion of all the tests, even if you are
aware that the analyzer did not pass. If there is a failure, complete the verification before
you troubleshoot, repair, and adjust. After the failure has been corrected, repeat the entire
system verification or performance tests and generate a new set of data.
Figure 3-1
ANSI/NCSL Z540–1–1994 Test Path Verification Flowchart
1. Stop only in case of a catastrophic failure or cable connector damage