Agilent Technologies E8362C Service Manual page 56

Pna series microwave network analyzers
Hide thumbs Also See for E8362C:
Table of Contents

Advertisement

Tests and Adjustments
System Verification
System Verification Procedure
1. Connect a cable to the analyzer as shown in
printer to the analyzer. For the printer, ensure that the correct driver is loaded and the printer is defined
as the default printer. Refer to the embedded help in the analyzer for printer setup. Let the analyzer warm
up for at least 90 minutes.
Figure 3-11
System Verification Test Setup
2. Insert the verification kit disk into the analyzer disk drive.
3. On the System menu, point to Service, and then click System Verification. The System Verification
dialog box is displayed; refer to
Figure 3-12
System Verification Dialog Box
3-24
Figure
3-11. If you desire printed test outputs, connect a
Figure
3-12.
PNA Series Microwave Network Analyzers
E8362C, E8363C, E836C
Service Guide E8364-90038