General Product Information

Required Service Test Equipment

Required Service Test Equipment
a
Critical Specifications
Equipment
Freq: 10 MHz to 20 GHz
Frequency counter
Accuracy : 0.5 ppm
Min Freq: 1 MHz
Spectrum analyzer
Max Freq: > 4 GHz
Resolution BW: 300 Hz
Power meter
Accuracy: ±0.0068 dB
Power sensor
Freq: 10 MHz to 4.2 GHz
(All)
Range: –30 to +20 dBm
Power sensor
Freq: 50 MHz to 20 GHz
(E8362C)
Range: –30 to +20 dBm
Power sensor
Freq: 50 MHz to 40 or 50 GHz
Range: –30 to +20 dBm
(E8363C, E8364C)
Dynamic accuracy
None specified
test set
Compression
None specified
test set
Digital voltmeter
Resolution: 10 mV
Printer
N/A
d
N/A
Test software
a. Unless specified otherwise, equipment listed is required for all analyzer models.
b. P = Performance tests, A = Adjustments, T = Troubleshooting, V = System verification
c. If an accurate measurement of the dynamic accuracy specification is not required, the E4418A or E4419A can be used.
d. The recommended model or part number for all equipment listed with a "P" in the Use column is required for proper operation of
this test software.
2-8
Recommended
Model or Part
Number
Test Instruments and Software
53151A
Opt 001
8565E
E4418B/19B
8482A
E4413A
8487A
Z5623A
Opt H01
Z5623A
Opt K01
Any
Any printer with Microsoft Windows 2000
driver
N7840A
PNA Series Microwave Network Analyzers
E8362C, E8363C, E8364C
Alternate Model
or Part Number
None
856xE
c
E4418A/19A
None
None
None
None
None
Any
None
Service Guide E8364-90038
b
Use
P, A,T
A,T
P, A,T
P, A,T
P, A,T
P, A,T
P
P
T