5.15.3 Specimen Stage Messages
Error Message
Please contact your local JEOL support center.
Error ID: 0200~0205
Please contact your local JEOL support center.
Error ID: 0240~0245
Please contact your local JEOL support center.
Error ID: 0261
Attempt to exceed the allowed stage range.
Error ID: 0230~0235
Stage stopped due to motor failure
Error ID: 0250~0255
Stage contact with lens detected.
Error ID: 0260
5.15.4 Messages about General Functions
Error Message
Please stop recording before quitting SEM control
software.
Please stop recording before switching the view
mode.
Stop recording now?
Recording failure.
The memory capacity might be insufficient or the
specified directory might be set to read-only mode.
Please stop recording before activating Halt Image.
Stop recording now?
Dead Time auto-set failure
Adjust PC manually.
During montage, SEM control software cannot be
exited
Finish montage and exit the SEM control software.
5.15.5 Messages about Options
Error Message
Please take a navigation image before stage control.
SNS image acquisition failure.
Readjustment required. Please contact your local
JEOL support center.
5-84
Cause & Response
Displayed when a motor driver (motherboard) overheats. The motors
for all axes become inoperable. Turn OFF the instrument power and
contact your local JEOL support center
Displayed when the mechanical movement limits are exceeded. Turn
OFF the instrument power and contact your local JEOL support center
Displayed when the R-axis continues to rotate for some reason. Turn
OFF the instrument power and contact your local JEOL support center
Displayed when the software movement limitations are exceeded.
Movements for all axes are stopped. Move back toward the origin (X,
Y) = (0, 0)
Displayed when the stage position cannot be read correctly.
Press the VENT button. After complete chamber venting execute stage
initialization (all axes).
Vent the specimen chamber to atmospheric pressure, and check
whether there is any damage to the specimen, objective lens, or
detectors.
Damage occurred...Contact your local JEOL support center.
No damage... Recheck the specimen shape and height
If the specimen protrudes above the specimen holder, be sure to
enter the amount of the protrusion (specimen height)
Cause & Response
Wait for the image recording to finish.
Click either the Yes or No button.
Check that the save destination exists, whether it has been set as
Read Only, and whether there is enough space available. Set a save
destination with adequate space.
Click either the Yes or No button.
Adjust gun alignment, and retry.
Wait for the montage to finish.
Cause & Response
Displayed when the click-center function is executed while the stage
navigation image is in the live image status.
There is no adjustment data for stage navigation control. Contact your
local JEOL support center.
No.ISMIT200-1E
Need help?