Instruments and instrument hoses

5.3.1

Purge the water lines (purge function)

To reduce bacteria, the water lines of the water-carrying instruments of
the dentist element and the Sprayvit multifunctional syringe of the
dentist and assistant element are purged with water.
Purge the water-carrying instruments after every patient with the purge
function before detaching the handpieces from the hose couplings for
hygienic reprocessing.
For the purge function, individual instruments are removed from their
holders and held over the cuspidor for purging. If your treatment center
is not equipped with a cuspidor, hold the instruments over a suitable
container. The water lines of all removed instruments are then purged
simultaneously. To purge the Sprayvit, its water button must be
pressed.
In the setup dialog, you can automatically block the user interfaces
when the purge function starts, see "Settings/Hygiene/Purge &
autopurge  [→ 178]/Purge with operation lock". The user interfaces can
then be wipe disinfected during or after flushing without inadvertently
triggering functions. Unblock them afterwards.
It is also possible to purge the water lines automatically, see "Purging
water lines automatically (auto-purge function)"  [→ 213].
Preparation
The following preparations should be made before you begin to purge
the water lines.
1. Set the quick couplings of the turbines to maximum water flow.
2. Place all instruments in the instrument holders. Ensure that all
instruments are placed in the correct holders. When the purge
program is active, instruments that are detected to be taken from
the holder are automatically purged, i.e. also instruments that were
inadvertently put in the wrong holder.
Open the "Purge" dialog
ü The start, instrument, or Sivision dialog is displayed.
1. Tap the Hygiene key in the footer to access the hygiene cockpit.
2. Tap on the "Purge" field.
Ä The "Purge" dialog appears.