5 Care, cleaning and maintenance by the practice team
Cleaning the drain lines of the cuspidor
The amalgam rotor is exposed to secretions, saliva and blood that
contain bacteria. It is therefore important to clean the drain lines before
replacing the amalgam rotor, see "Cleaning the drain lines of the
cuspidor"  [→ 243].
Removal and disposal of the amalgam rotor
WARNING
Wear gloves when performing the following work.
ü The treatment center is switched on.
1. Open the flap (A) on the base of the water unit.
2. Loosen the bayonet catch by turning the lower part of the amalgam
separator (B) counterclockwise. Remove the lower part of the
amalgam separator along with the amalgam rotor located inside it.
Ä A message appears on the Smart Touch and an acoustic signal
sounds.
3. Dispose of the amalgam residues from the cuspidor bowl and from
the suction hoses collected in the amalgam rotor; see "Cleaning the
gold trap"  [→ 241] and "Cleaning the filters in the suction hoses and
disinfecting the suction hoses"  [→ 239]. Put the amalgam residues
into the amalgam rotor.
4. Hold the lower part of the amalgam separator upright. Attach the
transport cap to the amalgam rotor.
Ä The transport cap latches in. Do not remove the transport cap
after closing it.
5. Remove the amalgam rotor with the transport cap from the lower
part of the amalgam separator.
6. Place the container in the special packaging and ship it for disposal
or authorize a certified waste management company.
Installing the amalgam rotor
Use only original Dentsply Sirona accessories. Never use a used or
recycled amalgam rotor.
To reorder the amalgam rotor, see "Spare parts and consumables" 
[→ 271].
1. Grease the O-ring (E) on the lower part of the amalgam separator.
For lubricants, see "Lubrication, cleaning, and disinfecting agents" 
[→ 198].
2. Insert the new amalgam rotor (C) in the lower part of the amalgam
separator (B).
5.5 Components of the water unit