Sterilization; Overview Of The Parts To Be Reprocessed - Dentsply Sirona Treatment Center Axano Pure Instructions For Use Manual

Table of Contents

Advertisement

Dentsply Sirona
Instructions for Use Treatment Center Axano Pure
Surfaces
Smart Touch user interface  [→ 202]
Dentist and assistant element as well as
plastic surfaces  [→ 201]
Removable parts of the dentist element 
[→ 204]
Silicone mat of the dentist and assistant
element  [→ 204]
Ball stopper  [→ 204]
Tray  [→ 206]
Cup holder  [→ 206]
Upholstery  [→ 207]
Handles of the LEDview Plus (see sepa-
rate LEDview Plus instructions for use)
Instruments
Housing, keyboard and nozzle of the
Sprayvit  [→ 219]
Motors  [→ 222]
68 29 514 D3685
D3685.201.02.01.02    2023-12
ISO 15883-1/-2, e.g. 93 °C (200 °F) and 10 min holding time. Refer to
the instructions for use of the device for the respective use. After the
process, check in good lighting (min. 500 lux) whether the product is
clean. If there is any contamination, repeat the process. If necessary,
perform sterilization afterwards.
5.1.2.3

Sterilization

Unwrapped sterilization
The products are reprocessed according to section "Cleaning and
disinfection"  [→ 196]. Use a steam sterilizer that complies with
EN 13060 Class B (e.g. DAC Premium / DAC Professional). Sterilize at
134 °C (274 °F), 3 minutes holding time and 1 minute drying time.
Wrapped sterilization
The products are reprocessed according to section "Cleaning and
disinfection"  [→ 196] and wrapped in packaging suitable for sterilization
and storage (paper/laminate packaging or ISO 11607 compliant
container). Use a steam sterilizer that complies with EN 13060 Class B
(e.g. DAC Premium / DAC Professional). Sterilize at 134 °C (274 °F), 3
minutes holding time and 1 minute drying time.
The specified holding time of 3 minutes is the minimum. Please observe
the relevant national regulations and recommendations.
5.1.2.4

Overview of the parts to be reprocessed

Cleaning and disinfection
Manual
Brush
(only
cleaning)
x
x
x
x
5 Care, cleaning and maintenance by the practice team
Automated
Wipe
Washer-
disinfec-
tor*
x
x
x
x
x
x
x
x
x
x
x
x
x
x
5.1 Basics
Sterilization
un-
wrapped
wrapped
x
x
x
x
x
197