Dentsply Sirona Treatment Center Axano Pure Instructions For Use Manual page 169

Table of Contents

Advertisement

Dentsply Sirona
Instructions for Use Treatment Center Axano Pure
68 29 514 D3685
D3685.201.02.01.02    2023-12
Water temperature
The temperature of the treatment water for the instruments on the
dentist and assistant element can be set in 5 increments. Check the
water temperature after filling a cup. The mixing temperature is in the
range of 35 °C to 45 °C.
WARNING
The heater power is adjustable.
Check the water temperature. The heater power should be reduced if
the water input is unusually warm.
After blow
After the foot control pedal has been released, the cooling spray
remaining in the instrument head or in the tip of the instrument can be
automatically blown out by briefly switching on the chip blower.
Camera button
The function of the camera key can be set: Delete last image, Focusing,
or Without function.
Export camera images
The photos and videos can be exported to Sidexis 4. You can set
whether the edited images should also be exported:
● Originals only
Only the original image made by the camera is exported.
● With changes
The original image and the edited image are exported.
● All
The original image, the edited image, and the edited image as a
scalable vector graphic (SVG) are exported.
4.16 Configuration of the treatment center (setup)
4 Operation
169