The NaCl pump has two settings for peristaltic pump hoses of different
diameters. For the Dentsply Sirona peristaltic pump hoses, the wide
setting must be selected. This ensures an optimal supply rate. The NaCl
pump is therefore delivered with the wide default factory setting.
The hose setting can be changed when needed by pressing on it with a
pen. The setting must always be adjusted on both sides.
4.5.8.8

Setting the NaCl flow rate

ü The expanded instrument dialog of the motor or ultrasonic
handpiece is displayed on the Smart Touch.
ü NaCl is preselected as the coolant, see "Preselecting coolant" 
[→ 87].
1. Use the arrow keys to set the flow rate of the NaCl pump.
Ä The selected setting is displayed on the upper edge of the
Smart Touch. In addition, the set flow rate is indicated by a gray
filling of the key.
2. Test the flow rate set for the instrument. Correct the setting if
necessary.
WARNING
Ultrasonic scaler tips of third-party manufacturers do not always have
a sufficient flow rate in conjunction with the NaCl function.
➢ Use only ultrasonic scaler tips from Dentsply Sirona. See also the
SiroSonic TL instructions for use.
4 Operation
4.5 Dentist element
91