Pre- And Postprocessing Of The Various Instruments - Dentsply Sirona DAC Universal Touch Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions DAC Universal
6681006 D3708
D3708.201.01.01.02    04.2019
4.2.4

Pre- and postprocessing of the various instruments

IMPORTANT
Please follow the regulations and guidelines applicable in your
country in your practice.
WARNING
Risk of contamination!
Please follow the instrument manufacturer's instructions when
handling the instruments.
(reprocessing instructions as per DIN EN ISO 17664 / -1)
NOTE
Please lubricate the collet system of your turbines, straight and
contra-angle handpieces regularly as per the lubrication and
operating instructions from the respective instrument manufacturer.
IMPORTANT
Reprocess immediately, or at the latest, one hour after treatment.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Predisinfect the instruments directly at the treatment center.
3. Transport the instrument to the hygiene room in a suitable transport
container.
4.2 Reprocessing with the DAC UNIVERSAL
4 Operation
69

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents