Step by Step Instructions
5 .3 .2

Attaching Handpiece Instruments

5 .3 .2 . 1 Attaching Electric Heat Plugger (EHP) to Pack Handpiece

Step
A
Select one of the three EHPs provided (small, medium or large).
B
Push the EHP into the end of the pack handpiece until it seats firmly.
EHPs can be installed in any one of 4 orientations.
To remove the EHP, wait for it to cool down, then pull straight out from hand-
C
piece.

5 .3 .2 .2 Attaching Thermal Response Tip (TRT) to Pack Handpiece

Step
A
Push the TRT into the end of the pack handpiece until it seats firmly.
TRTs can be installed in any one of 4 orientations.
To remove the TRT, wait for it to cool down, then pull straight out from hand-
B
piece.
The pack handpiece temperature must be set to 90°C when using the TRT to
prevent overheating and patient injury.