5 Care, cleaning and maintenance by the practice team
5.5 Components of the water unit
250
5.5.7

Changing the amalgam rotor

Amalgam residues and other solid particles are trapped in the amalgam
rotor according to the centrifugal principle.
If there are unread messages, a blue dot is displayed with the number
of unread messages on the setup key in the footer of the Smart Touch.
➢ Tap the adjacent symbol to open the "Notifications" dialog.
Ä If the amalgam rotor is nearly filled it must be replaced as soon
as possible. If the rotor is completely full, an acoustic signal
also sounds. A safety shutoff function ensures that the
treatment center cannot be used again until the rotor is
replaced.
CAUTION
Regardless of any message in Messages, the amalgam rotor must be
replaced at least once a year.
NOTICE
Amalgam residues must not enter the public sewage system.
Amalgam is a mercury compound that is hazardous to water.
➢ Do not dispose of amalgam residues in a sink.
➢ Collect amalgam residues in a closed container with water.
Dispose of amalgam residues, e.g. when replacing the amalgam
rotor by filling the amalgam residues into the amalgam rotor or
when emptying the sediment container.
WARNING
Disposal of the amalgam rotor
When a replacement rotor is supplied, a package for the return
shipment of the filled amalgam rotor is attached.
Authorize only certified waste management companies to dispose of
rotors.
Instructions for Use Treatment Center Axano Pure
D3685.201.02.01.02    2023-12
Dentsply Sirona
68 29 514 D3685